Frequency Accuracy
Measurement Lower Limit
(kHz)
Upper Limit
(kHz)
Measured Value
(kHz)
Pass/Fail
50 kHz
49.9985 50.0015
Anti-Alias Filter
Measurement Lower Limit Upper Limit
(dBfs)
Measured Value
(dBfs)
Pass/Fail
One Ch, Ch 1, 102.4 kHz
80
Two Ch, Ch 1, 51.2 kHz
80
Two Ch, Ch 2, 51.2 kHz
80
Input Coupling
Measurement Lower Limit Upper Limit
(dB)
Measured Value
(dB)
Pass/Fail
dc - ac, Ch 1
3
dc - ac, Ch 2
3
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Agilent 35670A Verifying Specifications
Performance Test Record - Two Channel
9of14