To change the fuses
Both fuses are installed at the factory.
•
Unplug the power cord from the analyzer.
•
Press in and turn the appropriate fuse holder cap counter-clockwise (use a
small screw driver for the ac fuse). Remove when the fuse cap is free from the
housing.
•
Pull the fuse from the fuse holder cap.
•
To reinstall, select the proper fuse and place in the fuse holder cap.
DC Fuse AC Fuse
HP 2110-0920 30 A 32 V Normal Blow HP 2110-0342 8 A 250 V Normal Blow
•
Place the fuse holder cap in the housing. Press in and turn clockwise.