To remove power supply
1 Remove rear panel (see ‘’To remove rear
panel’’).
2 Disconnect the ribbon cable from the A98
Power Supply assembly.
3 Using a T-15 torx driver, remove the six
screws from the A98 Power Supply assembly.
4 Set the front panel power switch in the
off (
O ) position (switch in the out position).
Replacing Assemblies Agilent 35670A
To remove power supply
6-14