To set up the anti-alias filter test
Performance Test only
This test verifies that the Agilent 35670A meets its FFT dynamic range
specification for frequency alias responses. In this test, a frequency synthesizer
outputs a 9 dBVrms signal known to cause an alias frequency to all channels. The
Agilent 35670A then measures the alias frequency to determine how well the alias
frequency was rejected. The test records at the end of this chapter list the
frequencies that are checked.
1 2ch
1 4ch
Agilent 35670A Verifying Specifications
To set up the anti-alias filter test
3-23