To set up the flatness test
Performance Test and Operation Verification
This test verifies that the Agilent 35670A meets its single channel amplitude
specification for FFT full scale flatness relative to 1 kHz. In this test, the ac
calibrator outputs a signal with an exact amplitude to all channels. The test records
at the end of this chapter list the amplitudes and frequencies that are checked.
1 2ch
1 4ch
Verifying Specifications Agilent 35670A
To set up the flatness test
3-18