Cross Channel Amplitude
FFT cross channel gain accuracy
Frequency response mode, same amplitude range
(AC coupled, Peroidic Chirp, Uniform Window, > =4Hz)
At full scale: Tested with 10 rms averages
on the −11 to +27 dBvrms ranges, and 100 rms
averages on the −51 dBVrms range
±0.04 dB (0.46%)
At −20 dBfs: Tested with 200 rms averages on
the −11 to +27 dBVrms ranges, and 2000 rms
averages on the −51 dBVrms range
±0.08 dB (0.92%)
Cross Channel Phase
Cross channel phase accuracy
(same conditions as cross-channel amplitude
>=12Hz)
±0.5 degree