Routine Maintenance: Inlets
Maintaining a purged packed inlet
Released: March 2004 6850 Series II GC User Information page 136 of 256
Table 27. Liners and Inserts for Packed Columns
Column type Liner Insert
1/8-in. metal 1/8-inch stainless steel
19243-80510
None
19243-80530 5080-8732 or
5181-3382*
1/4-inch metal 1/4-inch stainless steel
19243-80520
None
19243-80540 5080-8732 or
5181-3382*
1/4-inch glass No liner required.
Column end functions
as liner. Can also use
1/4-inch metal liner.
Not applicable
*Deactivated
1/4- or 1/8-inch liner
(stainless steel)
Insert
1/4-inch Vespel ferrule
1/4-inch nut
1/8-inch or 1/4-inch
stainless steel liner