EasyManuals Logo

Agilent Technologies 6850 Series II User Manual

Agilent Technologies 6850 Series II
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
Routine Maintenance: Inlets
Maintaining a purged packed inlet
Released: March 2004 6850 Series II GC User Information page 136 of 256
Table 27. Liners and Inserts for Packed Columns
Column type Liner Insert
1/8-in. metal 1/8-inch stainless steel
19243-80510
None
19243-80530 5080-8732 or
5181-3382*
1/4-inch metal 1/4-inch stainless steel
19243-80520
None
19243-80540 5080-8732 or
5181-3382*
1/4-inch glass No liner required.
Column end functions
as liner. Can also use
1/4-inch metal liner.
Not applicable
*Deactivated
1/4- or 1/8-inch liner
(stainless steel)
Insert
1/4-inch Vespel ferrule
1/4-inch nut
1/8-inch or 1/4-inch
stainless steel liner

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals