EasyManuals Logo

Agilent Technologies 6850 Series II User Manual

Agilent Technologies 6850 Series II
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
Routine Maintenance: Inlets
Maintaining a PTV
Released: March 2004 6850 Series II GC User Information page 154 of 256
Replacing the Teflon ferrule
1. Load the SERVICE method or reset the inlet and oven temperatures to
OFF or turn the main power switch off. Let the heated zones cool to room
temperature. Also, make sure that:
If you have entered parameters that you do not want to lose, store
them as a method
If the detector is on, turn it off.
2. Unscrew the guide cap from the septumless head and remove the Teflon
ferrule.
3. Push the guide cap and the new Teflon ferrule over the syringe needle so
that at least 10 mm of the needle tip is exposed.
4. Guide the end of the syringe needle into the septumless head until the
ferrule meets the septumless head.
5. Tighten the guide cap until resistance is first felt.
6. Check for leaks when the syringe needle has been fully introduced.
7. If necessary, carefully tighten the guide cap until the inlet stops leaking.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series II and is the answer not in the manual?

Agilent Technologies 6850 Series II Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series II
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals