Automatic Liquid Sampler and Injector
The 6850 automatic liquid sampler
Released: March 2004 6850 Series II GC User Information page 231 of 256
Filling the turret for use
1. Load all samples to be run into the turret. Make sure that you place them
in the turret positions that correspond to the sequence (Control Module or
ChemStation/Cerity Chemical control) or Work List (Cerity control). The
vial positions are labeled.
2. Load clean, empty waste bottles into the appropriate waste locations.
Always load a bottle in the Waste C location.
3. Load your solvent bottles into the solvent locations as needed.
Removing the turret
1. Remove all vials from the turret.
2. Remove the syringe and the needle support foot.
3. Remove the two plastic screws in the ESD cap, and remove the cap from
the motor hub.
Caution Do not loosen the central mounting screw. This screw secures the motor hub in
place.
ESD cap
Remove screws