Messages
Fault messages
Released: March 2004 6850 Series II GC User Information page 97 of 256
FAULT 242 Aux3 faulty fact cal
FAULT 243 Aux4 faulty fact cal
FAULT 244 Aux5 faulty fact cal
Possible cause: Defective auxiliary flow module
The auxiliary flow module’s calibration is invalid.
FAULT 245 Detector module rev
FAULT 247 Inlet module rev
FAULT 249 Aux module revision
Possible cause: Flow module version is obsolete and is not compatible with the GC
The flow module’s EEPROM is not recognized by the GC. The module is
unusable. The version of the module is not compatible with the version of the
GC.
FAULT 250 Wrong det module
FAULT 252 Wrong inlet module
FAULT 254 Wrong aux module
Possible causes: Wrong flow module installed for detector/inlet/aux
Defective detector/inlet/aux flow module
Flow module revision is obsolete and not recognized
The flow module installed for the detector/inlet/aux is unusable. The module is
not recognized as a valid detector flow module.
FAULT 255 Det invalid type
FAULT 257 Inlet invalid type
Possible cause: The installed detector/inlet flow module is not supported by the GC
The detector/inlet flow module revision is not supported by the firmware
installed in the GC.
FAULT 259 Det type mismatch
Possible cause: GC detector type was changed but its board was not replaced or the reverse
The detector flow module type does not match the installed detector
electronics board.
FAULT 262 Front RS232 failed
FAULT 263 Back RS232 failed
FAULT 264 Sampler RS232 failed
Possible causes: Defective main board
The RS232 communications on the main board are faulty. Communication
between the GC and the other device is halted.