100 Operation Manual
5 Using the Removable Ion Source
To Remove the Source from the Analyzer Chamber
2 Keeping the RIS handle in position, push the RIS handle
gently and allow the RIS bayonet to extend into the analyzer
chamber. Make sure the RIS handle extends fully, to ensure
that the RIS bayonet seats in the holes on the ion source body
properly.
3 When the ion source is properly seated, turn the RIS handle
counter clockwise until the alignment guide is in the twelve
o’clock position, to lock the ion source in place.
4 Pull the RIS handle toward you, keeping the alignment guide
at the twelve o’clock position, to retract the ion source into
the RIS cooling chamber.
5 Return to step 8 in the “To Change the Ion Source” on
page 94.