102 Operation Manual
5 Using the Removable Ion Source
To Change the Ion Source on the RIS Probe Extraction Tool
9 Guide the pin on the RIS bayonet into the tracks on the end
of the ion source body.
10 While holding the ion source securely, turn the RIS handle
counter clockwise so that the alignment guide is in the twelve
o’clock position, to lock the ion source securely on the
bayonet.
11 Retract the RIS handle fully so that the ion source is located
completely in the cooling chamber.
12 Raise the locking pin on the RIS cooling chamber hinge and
swing it shut.
13 Secure the RIS cooling chamber to the instrument with the
latch on the right side of the door.
14 Return to step 10 in the “To Change the Ion Source” on
page 94.