Operating in Chemical Ionization (CI) Mode 4
To Perform a CI Autotune
Operation Manual 89
8 Select the Tune from default settings check box if you are
restarting the system after a system vent, major servicing, or
a power outage. If you clear the Tune from default settings box,
the Autotune process starts using previous tune values.
9 Select Save tune file when done check box to save the new tune
parameters generated by the autotune. Do not select this
item if you want to review the autotune report before saving
the newly generated tune parameters.
10 Select the Print autotune report check box to automatically
print a tune report.
11 Click the Autotune button to start the autotune. The Status
line displays the current step in the autotune process and the
plot of the tuned parameter for that step is shown in the top
graph. If specified above, at the completion of the autotune a
Tune Report is printed.
To stop the autotune before it completes the autotune
parameter selection, click the Abort Autotune button. The
parameters from the last successful autotune are used.
12 Review the Tune Report. If the results are acceptable and you
did not select the Save tune file when done check box, save the
autotune by clicking the Files and Report tab, then click the
Save button.

Table of Contents

Other manuals for Agilent Technologies 7200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7200 and is the answer not in the manual?

Agilent Technologies 7200 Specifications

General IconGeneral
BrandAgilent Technologies
Model7200
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals