Agilent 1260 Infinity RID User Manual 125
Test Functions
8
Optical Balance
3 Adjust optical balance.
a While monitoring the optical balance use the flat-headed screwdriver to
turn the zero glass adjustment screw slowly (see Figure 34 on page 125).
b When the diode balance value reaches 0.00 optical balance is restored.
Figure 34 Turning the zero glass adjustment screw
OZgd\aVhhVY_jhibZci
hXgZlaZ[i

Table of Contents