EasyManua.ls Logo

Agilent Technologies Infinity 1260 - Page 74

Agilent Technologies Infinity 1260
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
74 Agilent 1260 Infinity RID User Manual
4 Using the Refractive Index Detector
Running a Checkout Sample
4 Turn the heater ON and purge the detector reference cell for 20 minutes as
shown in Figure 26 on page 74:
Figure 26 RID Check Out Sample Control
5 When purging has finished allow the baseline to stabilize and start the
analysis.

Table of Contents

Related product manuals