Agilent 1290 Infinity II Flexible Pump User Manual 113
Maintenance
7
Pump Head Procedures
7
To remove wear, use isopropanol.
8 To clean the piston and the pump head sufrace, use the
ultra fine abrasive mesh (p/n 8660-0852).
9 To dry the pump head and to remove residues, use
pressurized air ((1), (2) and (3)).
10 If the wash seal shall be replaced by a new one, use the
steel side of the insert tool for removing it.
&
(
'
Wash seal