Agilent 1290 Infinity II Flexible Pump User Manual 27
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tabl e 1 Physical Specifications
Type Specification Comments
Weight 16.1 kg (35.5 lbs)
Dimensions
(height × width × depth)
180 x 396 x 436 mm
(7.1 x 15.6 x 17.2 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 120 VA / 110 W
Ambient operating
temperature
4 – 55 °C (39 – 131 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.