EasyManua.ls Logo

Agilent Technologies Infinity II 1290

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1290 Infinity II Flexible Pump User Manual 27
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tabl e 1 Physical Specifications
Type Specification Comments
Weight 16.1 kg (35.5 lbs)
Dimensions
(height × width × depth)
180 x 396 x 436 mm
(7.1 x 15.6 x 17.2 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 120 VA / 110 W
Ambient operating
temperature
4 – 55 °C (39 – 131 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.

Table of Contents

Related product manuals