EasyManuals Logo

Agilent Technologies Infinity II 1290 User Manual

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
Agilent 1290 Infinity II Flexible Pump User Manual 63
Error Information
6
Pump Error Messages
Quaternary pump shutdown during analysis
Error ID: 29199
The quaternary pump has been shut down by the control software or
control module during an analysis.
Parameter: 0 for off, 1 for standby.
Reading the pump encoder tag failed
Error ID: 29201
Reading the pump encoder tag has failed.
Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Pump has been shut down.
Restart pump.
Probable cause Suggested actions
1 Defect connection between encoder and
main board.
Please contact your Agilent service
representative.
2 Missing or defect tag Defect connection
between tag and encoder.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals