36 Agilent 1290 Infinity II Flexible Pump User Manual
3 Using the Pump
Best Practices
Prepare the pump
Purge
Use the Purge function to:
fill the pump,
exchange a solvent,
remove air bubbles in tubes and pump heads.
Condition
Use the Conditioning function:
daily when starting the pump,
to minimize pressure ripple by dissolving air bubbles in pump heads.
Seal wash
The seal wash function runs continuously and is controlled by the seal
wash sensor. This guarantees a maximum seal life time.
NOTE
Condition your complete system with solvents and composition of your application (for
example 50 %/50 % A/B at 0.5 mL/min.
CAUTION
Contaminated seal wash solvent
Do not recycle seal wash solvent to avoid contamination.
Weekly exchange seal wash solvent.