EasyManua.ls Logo

Agilent Technologies Infinity II 1290

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 Agilent 1290 Infinity II Flexible Pump User Manual
6 Error Information
Pump Error Messages
Pump Error Messages
These errors are pump specific.
Pressure of quaternary pump above upper limit
Error ID: 29163
The pressure has exceeded the upper pressure limit.
Parameter: Measured pressure
Pressure below lower limit
Error ID: 29176
The pressure has dropped below the lower limit.
Parameter: None
Probable cause Suggested actions
1 Blockage in flow path after the pressure
sensor.
Check for blockages in the LC system, e.g. purge
valve, Jet Weaver, degraded column, column frits,
needle, needle seat, capillaries etc.
Check for particles in the solvent.
2 Inappropriate settings (pressure limit, flow
rate).
Decrease flow rate.
Increase pressure limit.
Probable cause Suggested actions
1 Leak
Check for leaks.
2 Bottle empty
Check bottle filling.
3 Wrong solvent (viscosity)
Check solvent.
4 Inappropriate setting
Check flow rate and lower pressure limit.
5 Column degradation
Replace column.

Table of Contents

Related product manuals