EasyManua.ls Logo

Agilent Technologies Infinity II 1290

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
78 Agilent 1290 Infinity II Flexible Pump User Manual
6 Error Information
Pump Error Messages
Valve hardware overcurrent (MCGV)
Error ID: 29227
Power consumption too high for one of the MCGV valves.
Probable cause Suggested actions
1 Cable defect.
Replace MCGV.
2 Valve defect
Replace MCGV.
3 Defective main board.
Please contact your Agilent service
representative.