104 Agilent 1290 Infinity II Flexible Pump User Manual
7 Maintenance
Pump Head Procedures
Pump Head Procedures
CAUTION
Limitation of life time
The pump head assembly is an exchange part which cannot be reassembled with
standard tools. Disassembling the pump head will strongly limit its life time.
Do not disassemble the pump head assembly.
CAUTION
Damage of connections
Disassembling the flow connection between the two pump heads of the pump head
assembly (solvent channel) can damage the connection and cause leaks.
Do not disconnect the flow connection between the pump heads.
CAUTION
Damage of internal parts
Do not apply a strong force to the screws of the pump head.
Use a torque hex key for that purpose.
CAUTION
Damage of the pump piston
Removing pump heads in a position other than the maintenance position can damage
the pump piston.
Before switching off the pump, bring it to the maintenance position.
CAUTION
Damage of pump drives
The pump drive can be damaged if the pump initializes after switching it on without
having the pump head installed properly.
Use the Lab Advisor maintenance procedure for replacing pump heads.
Install the pump head correctly before switching on the pump.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals