72 Agilent 1290 Infinity II Flexible Pump User Manual
6 Error Information
Pump Error Messages
Pump drive error
Error ID: 29212, 29213
The pump drive failed during calibration.
Parameter: 1 – 2 referring to pump drive
Maximum stroke is too short
Error ID: 29203
The maximum stroke is too short.
During initialization the pump defines the operation position of the pump
drives and therefore the pistons. First the pump drive moves backwards to
find a mechanical stop within the ball screw. Afterwards, pistons move
forwards for finding the maximum available stroke volume. These values
are expected within a pre- defined range. "Maximum stroke too short"
means that the outer drive position is too close. This can be caused by a
drive initialization without pump head or if the pump head has not been
installed properly (screws are loose).
Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Wiper shifted
Please contact your Agilent service
representative.
2 Pump head blocks piston movement
Replace, clean or repair pump head.
3 Pump drive motor is mechanically blocked.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals