58 Agilent 1290 Infinity II Flexible Pump User Manual
6 Error Information
General Error Messages
Compensation Sensor Short
Error ID: 0080
The ambient- compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
power switch board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient
temperature changes. If the resistance across the sensor falls below the
lower limit, the error message is generated.
Probable cause Suggested actions
1 Defective power switch board
Please contact your Agilent service
representative.
2 Loose connection between the power
switch board and the main board
Please contact your Agilent service
representative.