Agilent 1290 Infinity II Flexible Pump User Manual 35
Using the Pump
3
Best Practices
Best Practices
Daily / Weekly tasks
Daily tasks
Replace mobile phase based on water/buffer.
Replace organic mobile phase latest every 2nd day.
Check seal wash solvent.
Run conditioning with composition of your application.
Weekly tasks
Change seal wash solvent (10 % / 90 % isopropanol/water) and bottle.
Flush all channels with water to remove salt deposits.
Visually inspect solvent filters. Clean or exchange if necessary.
Power up / Shut-down the pump
Power up the pump
Use new or different mobile phase (as required).
Purge pump heads with 2.5 – 3 mL/min for 5 min.
Condition pump heads for 10 – 20 min.
Long-term shut-down of the pump
Flush system with water to remove buffer.
Use recommended solvents to store the system.
Power off the pump or system.