EasyManuals Logo

Agilent Technologies Infinity II 1290 User Manual

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
66 Agilent 1290 Infinity II Flexible Pump User Manual
6 Error Information
Pump Error Messages
Drive timeout
Error ID: 29204
Movement of drive during initialization is blocked mechanically.
Parameter: 1 – 2 referring to pump drive
Overcurrent of pump drive
Error ID: 29202
The current consumption of the pump drive is too high.
Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Blockage in flow path
Remove capillary connection to system, check
outlet filter, check valves, check pump head.
2 Blockage of pump drive Drive motor defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Blockage of system before pressure sensor.
Check for blockage of e.g. outlet valve filter frit,
Multi Purpose Valve, heat exchanger.
2 Drive motor defect.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals