Agilent 1290 Infinity II Flexible Pump User Manual 37
Using the Pump
3
Best Practices
How to deal with solvents
Use clean bottles only.
Exchange water- based solvents daily.
Select solvent volume to be used up within 1 – 2 days.
Use only HPLC- grade solvents and water filtered through 0.2 µm filters.
Label bottles correctly with bottle content, and filling date / expiry
date.
Use solvent inlet filters.
Reduce risk of algae growth: use brown bottles for aqueous solvents,
avoid direct sunlight.
Select channels for Multi-Channel Gradient Valve (MCGV)
Use lower channels (A and/or D) for buffer solutions.
Regularly flush all MCGV channels with water to remove possible salt
deposits.
Check compatibility of buffers and organic solvents to avoid
precipitation.