Maintenance
7
Pump Head Procedures
Replacing the Heat Exchanger
Tools required Description
Wrench, 19 mm
Parts required p/n Description
G4220-81013 Heat Exchanger (secondary pump head only)
G4220-20028 Headless screw for 1290 Infinity pump heads
G4220-20001 Spacer Fitting
Preparations Remove the pump head assembly from the pump
Remove the secondary pump head from the link plate
CAUTION
Loss of small spacer fitting
Inside the secondary pump head is a small spacer fitting, which can be dropped easily
when removing the heat exchanger.
The heat exchanger does not need to be removed for pump head maintenance.
1 Remove the 19 mm screw at the front of the secondary
pump head.
2 Remove the front plate.