EasyManuals Logo

Agilent Technologies Infinity II 1290 User Manual

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
Agilent 1290 Infinity II Flexible Pump User Manual 117
Maintenance
7
Pump Head Procedures
Replacing the Heat Exchanger
Tools required Description
Wrench, 19 mm
Parts required p/n Description
G4220-81013 Heat Exchanger (secondary pump head only)
G4220-20028 Headless screw for 1290 Infinity pump heads
G4220-20001 Spacer Fitting
Preparations Remove the pump head assembly from the pump
Remove the secondary pump head from the link plate
CAUTION
Loss of small spacer fitting
Inside the secondary pump head is a small spacer fitting, which can be dropped easily
when removing the heat exchanger.
The heat exchanger does not need to be removed for pump head maintenance.
1 Remove the 19 mm screw at the front of the secondary
pump head.
2 Remove the front plate.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals