EasyManuals Logo

Agilent Technologies Infinity II 1290 User Manual

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
62 Agilent 1290 Infinity II Flexible Pump User Manual
6 Error Information
Pump Error Messages
Waste counter limit exceeded
Error ID: 29147
The counter for the waste volume has exceeded the limit, which has been
set in the user interface.
Parameter: None
Flow rate limit exceeded
Error ID: 29164
The flow rate of the quaternary pump has exceeded the limit, while the
pump runs in pressure controlled mode, e.g. during a pressure test.
Parameter: None
Probable cause Suggested actions
1 The waste container is full.
Empty waste container.
2 Inappropriate setting for waste counter.
Reset waste counter.
Adjust waste counter limit.
Probable cause Suggested actions
1 Leak
Check for leaks in the pump and flow path.
2 Bottle empty.
Fill solvent bottle.
3 Shutoff valve closed (if applicable).
Open shutoff valve.
4 Drift of pressure sensor (unlikely for short
tests taking some minutes).
Replace pressure sensor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals