Error Information
6
Pump Error Messages
Degasser's pressure limit violation
Error ID: 29220
Pressure too far above the limit.
Seal wash pump was missing when tried to turn on
Error ID: 29223
The seal wash pump has not been detected (while being configured or
detected before)
Probable cause Suggested actions
1 Leak in degasser chamber or degasser
tubing.
Please contact your Agilent service
representative.
2 Defect vacuum pump.
Please contact your Agilent service
representative.
3 Degasser chamber empty or connected to
air.
Block unused degasser channels.
Probable cause Suggested actions
1 Defect cable connection to seal wash pump.
Check cable connection.
2 Defect seal wash pump motor.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals