Agilent 1290 Infinity II Flexible Pump User Manual 77
Error Information
6
Pump Error Messages
Degasser's pressure limit violation
Error ID: 29220
Pressure too far above the limit.
Seal wash pump was missing when tried to turn on
Error ID: 29223
The seal wash pump has not been detected (while being configured or
detected before)
Probable cause Suggested actions
1 Leak in degasser chamber or degasser
tubing.
Please contact your Agilent service
representative.
2 Defect vacuum pump.
Please contact your Agilent service
representative.
3 Degasser chamber empty or connected to
air.
Block unused degasser channels.
Probable cause Suggested actions
1 Defect cable connection to seal wash pump.
Check cable connection.
2 Defect seal wash pump motor.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.