Agilent 1290 Infinity II Flexible Pump User Manual 61
Error Information
6
Pump Error Messages
Target pressure not reached for quaternary pump degasser
Error ID: 29221
The target pressure of the quaternary pump degasser has not been
reached within the expected time.
Parameter: Pressure in mbar
Solvent counter exceeded limit
Error ID: 29146
The counter for the solvent volume has exceeded the limit, which has been
set in the user interface.
Probable cause Suggested actions
1 Condensation in degasser chamber due to
temperature fluctuation.
Equilibrate and restart module.
2 Degasser is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 No solvent present.
Refill solvent bottle.
2 Inappropriate setting.
Check solvent counter setting in user interface.