76 Agilent 1290 Infinity II Flexible Pump User Manual
6 Error Information
Pump Error Messages
Drive Encoder failed
Error ID: 29210
Drive encoder failed during pump drive calibration.
Drive phases differ too much in electric resistance
Error ID: 29216
Pump drive calibration has failed due to a strong difference electric
resistances for different motor phases.
Probable cause Suggested actions
1 Internal error.
Contact Agilent support.
Probable cause Suggested actions
1 Pump drive cable defect.
Please contact your Agilent service
representative.
2 Pump drive defect.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.