Agilent 1290 Infinity II Flexible Pump User Manual 71
Error Information
6
Pump Error Messages
Writing the multi purpose valve tag failed
Error ID: 29241
Writing the multi purpose valve tag failed.
Parameter: None
Unknown multi purpose valve type
Error ID: 29242
The type information of the multi purpose valve is invalid.
Parameter: None
Pump drive encoder error
Error ID: 29211
The pump drive encoder has generated no signal.
Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Multi purpose valve head tag defect.
Replace multi purpose valve head.
2 Multi purpose valve tag head reader is
defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Wrong valve head installed.
Check or replace multi purpose valve head.
2 Valve head has invalid RFID tag content.
Check or replace multi purpose valve head.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infinity II 1290 and is the answer not in the manual?

Agilent Technologies Infinity II 1290 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinity II 1290
CategoryWater Pump
LanguageEnglish

Related product manuals