Agilent 1290 Infinity II Flexible Pump User Manual 69
Error Information
6
Pump Error Messages
Reading of multi purpose valve tag failed
Error ID: 29240
Reading the multi purpose valve tag failed.
• Parameter: None
Pump drive encoder rollover
Error ID: 29232
Invalid pump drive encoder signals have been detected.
• Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Reading of multi purpose valve tag failed.
Check cable connection.
2 Multi purpose valve head tag defect or
empty.
Replace multi purpose valve head.
3 Multi purpose valve tag reader is defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.