EasyManuals Logo

Agilent Technologies Infinity II 1290 User Manual

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #67 background imageLoading...
Page #67 background image
Agilent 1290 Infinity II Flexible Pump User Manual 67
Error Information
6
Pump Error Messages
Deliver underrun
Error ID: 29233
Internal error.
Parameter: None
Defect connection between main board and pump drive encoder
Error ID: 29208
Defect connection between main board and pump drive encoder.
Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Internal error.
Please contact your Agilent service
representative.
2 Firmware issue
Use a minimum firmware revision of B.06.55
Probable cause Suggested actions
1 Defect connection between main board and
pump drive encoder.
Please contact your Agilent service
representative.
2 Defect encoder.
Please contact your Agilent service
representative.