132 Agilent 1290 Infinity II Flexible Pump User Manual
7 Maintenance
Pump Head Procedures
3
Insert the new inlet weaver to the inlet valve (1, 2). Fix
the weaver with the plastic screw to the inlet valve (3).
4 Reconnect solvent tubes for channels A-D from the
MCGV to the degasser outlets.
5 Connect the capillary from the outlet filter adapter on the
secondary pump head to the pressure sensor.
6 Connect the seal wash tubes.
Next Steps:
7 Close the doors.
8 Perform a Pump Leak Rate Test and a System Pressure Test.