EasyManua.ls Logo

Agilent Technologies Infinity II 1290

Agilent Technologies Infinity II 1290
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
132 Agilent 1290 Infinity II Flexible Pump User Manual
7 Maintenance
Pump Head Procedures
3
Insert the new inlet weaver to the inlet valve (1, 2). Fix
the weaver with the plastic screw to the inlet valve (3).
4 Reconnect solvent tubes for channels A-D from the
MCGV to the degasser outlets.
5 Connect the capillary from the outlet filter adapter on the
secondary pump head to the pressure sensor.
6 Connect the seal wash tubes.
Next Steps:
7 Close the doors.
8 Perform a Pump Leak Rate Test and a System Pressure Test.
8dccZXi^dcid^caZilZVkZg
&
'
(

Table of Contents

Related product manuals