Agilent 1290 Infinity II Flexible Pump User Manual 137
Maintenance
7
Replace the High Pressure Outlet Filter or Filter Frit
3
Unscrew the high pressure outlet filter. 4 Replace the filter frit.
5 Mount the high pressure outlet filter using a torque
wrench set to 16 Nm.
6 Reinstall the high pressure outlet filter to the filter holder.
7 Mount the capillary connection to the pressure sensor (1)
and to the Multipurpose valve (2).
;g^i
'#
&#