Agilent 1290 Infinity II Flexible Pump User Manual 139
Maintenance
7
Replace Parts of the Inline Filter
3
Use two 5/16 wrenches for opening the inline filter. 4 Replace the filter frit and reassemble the inline filter.
5 Put the inline filter to the clamp and install its capillaries.
The integrated capillary is connected to port 5 of the
Multi Purpose Valve. The removable capillary is
connected to port 8.