Agilent 1290 Infinity II Flexible Pump User Manual 91
Maintenance
7
Replace the Inlet Weaver
3
Pull the Inlet Weaver out of the Inlet Valve. 4 Insert the new inlet weaver to the inlet valve. Fix the
weaver with the plastic screw.
5 Fix the fitting of the new inlet weaver to the MCGV. 6 Reconnect tubings between MCGV and degasser.