PL-GPC 220 User Manual 65
Error Information
6
Errors
The error will also be written to the error log file, which can be viewed by
selecting the View Error log option from the Log menu.
Tab le 3 Error overview
Type GUI Err or
Message
Conditions Response Time
(Timeout, Retries)
LED Beeper Default
Mode
Notes
Pump
PRESSURE "Pressure is
currently outside
set limits"
Pressure has
exceeded the
set min or
max pressure
settings
10 s (10 s, 1) Flash
Orange
NO IDLE Not STBY
and not
initializing
50 s (10 s, 5) Flash
Red
YES STBY "
PUMP COMMS "The pump is no
longer
responding"
No serial
response from
the pump
3 s (1 s, 3) Flash
Red
YES STBY
Oven
TEMPERATURE "Oven
temperature has
become unstable"
Tem p + /- 2 ° C
from setpoint
- (-, -) Flash
Red
YES STBY Not STBY
FAN "The fan is no
longer operating
correctly"
Frequency <
100 Hz
5 s (5 s) Flash
Red
YES STBY "
DOOR "Illegal opening of
the oven door has
occurred"
I/O state
change
- (-, -) Flash
Orange
YES Heaters
OFF while
door open
V/F "Temperature
sensor is not
responding
correctly"
Tem p = 0 ° C or
> 999.99 °C
- (-, -) Flash
Red
YES STBY "
SENSOR "The vapour
sensor is not
responding
correctly"
Open/Short
circuit
5 s (5 s) Flash
Red
YES STBY "

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PL-GPC 220 and is the answer not in the manual?

Agilent Technologies PL-GPC 220 Specifications

General IconGeneral
BrandAgilent Technologies
ModelPL-GPC 220
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals