66 PL-GPC 220 User Manual
6 Error Information
Errors
OVEN VAPOUR "Spillage may
have occurred
within the oven"
Voltage < set
threshold
2 min (-, -) Flash
Orange
YES STBY Immediat
e alarm
5 min (-, -) " YES - "
RI
AZ FAIL "Attempts to
autozero the RI
have failed"
Timeout or
False Zero
20 min (10 min, 2) Flash
Orange
NO - Not STBY
and not
initializing
PURGE "The purge valve is
not functioning
correctly"
Timeout 5 s (5 s) Flash
Red
NO IDLE "
INITIALIZATION "Failed to initialize
autosampler"
cmd failed
response from
autosampler
- (-, 0) Flash
Red
NO -
MOVE "Failed to move
carousel correctly"
" - (-, 3) Flash
Red
NO -
VIAL LOWER "Failed to lower
vial from the oven"
" - (-, 0) Flash
Red
NO -
VIAL LIFT "Failed to lift vial
from the carousel"
" '- (-, 0) Flash
Red
NO -
SYRINGE "Failed to move
syringe"
" - (-, 0) Flash
Red
NO -
CAROUSEL LOCK "Failed to lock
carousel after
being closed"
cmd failure
during
autosampler
run
- (-, 0) Flash
Red
NO -
LOAD "Failed to switch
valve to load
position"
cmd failed
response from
autosampler
- (-, 3) Flash
Red
NO -
INJ "Failed to switch
valve to inject
position"
" - (-, 3) Flash
Red
NO -
Tab le 3 Error overview
Type GUI Err or
Message
Conditions Response Time
(Timeout, Retries)
LED Beeper Default
Mode
Notes