Page 34 VK 7025 Dissolution Apparatus Revision H, 11/10
Setup Operator’s Manual P/N 70-9033
Varian, Inc.
Centering Verification
To verify the centering of the paddle / basket shafts, complete the following steps:
Step 1. Locate the centering verification gauge and the EaseAlign centering ring.
FIGURE 6. Centering Verification Gauge
FIGURE 7. EaseAlign Centering Ring
Step 2. Place the EaseAlign centering ring on the first vessel (see Figure 7,
“EaseAlign Centering Ring,” above).
Step 3. Place the centering verification gauge against the shaft with the shaft
resting in the radial cutout (see Figure 6, “Centering Verification Gauge,”
above).
radial cutout