Revision H, 11/10 VK 7025 Dissolution Apparatus Page 57
P/N 70-9033 Operator’s Manual Administrator Operation
Varian, Inc.
Press ESC to return to the System Setup Menu 2 screen.
Setting Serial Numbers
To enter the serial numbers of your accessories, complete the following steps:
Step 1. Select SERIAL NUMBERS from the System Setup Menu 2 screen. The
Enter Serial Numbers screen displays.
Step 2. Select VESSEL, PADDLE, BASKET / SHAFT, or ROTATING
CYLINDER. The corresponding serial number screen displays with
numbers 1 through 8 listed down the left side of the screen. If BASKET /
SHAFT is selected, two columns display. The left column is for basket
serial numbers and the right column is for shaft serial numbers.
Baud Rate Select BAUD RATE to toggle between 2400, 4800, 9600, 19200, 28800,
and 38400. The baud rate must be set to 9600 when connected to a
VK 8000 or Total Solution system.
Comm. ID Select COMM. ID. The cursor flashes. Enter the correct communications
port identification number in xx format and press ENTER. The correct
communications port identification number displays. The
communications port identification number must be 01 when connected
to a VK 8000 or single Total Solution system.
Note: a leading zero is necessary when entering a communications
port identification number between 01 and 09.
Option Response

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Varian VK 7025 and is the answer not in the manual?

Agilent Technologies Varian VK 7025 Specifications

General IconGeneral
BrandAgilent Technologies
ModelVarian VK 7025
CategoryMedical Equipment
LanguageEnglish

Related product manuals