Revision H, 11/10 VK 7025 Dissolution Apparatus Page 53
P/N 70-9033 Operator’s Manual Administrator Operation
Varian, Inc.
Calibration Calendar
From the Setup Alarms screen, select USP CAL MENU. The Calibration Calendar screen
displays.
Following is a description of the Calibration Calendar screen options:
An alarm sounds when an entered date is reached.
ALARM BASKET CAL DUE, ALARM PADDLE
CAL DUE, or ALARM PM DUE displays. Press CLEAR to silence the alarm.
Press ESC to return to the Setup Alarms screen. When all appropriate parameters on the
Setup Alarms screen have been entered, press ESC to return to the System Setup Menu
1 screen.
Option Response
Basket Calibration Date Enter the date the calibration was performed in the
appropriate format and press ENTER.
Paddle Calibration Date Enter the date the calibration was performed in the
appropriate format and press ENTER.
Basket Expiration Date Enter the calibration expiration date in the
appropriate format and press ENTER.
Paddle Expiration Date Enter the calibration expiration date in the
appropriate format and press ENTER.