Page 98 VK 7025 Dissolution Apparatus Revision H, 11/10
Maintenance and Troubleshooting Operator’s Manual P/N 70-9033
Varian, Inc.
thoroughly after use. If any stainless steel parts show signs of surface
discoloration, lightly wipe the surface with a soft cloth or nonabrasive pad to
remove it.
Carefully wipe the paddles or basket shafts after each use.
Carefully wash the dissolution baskets after each use.
Carefully wash the dissolution vessels after each use.
Weekly Maintenance
See “Paddle / Basket Shaft Care” on page 99, “Basket Care” on page 100, and “Water
Bath / Acrylic Care” on page 101 as applicable for additional information on proper
maintenance of your equipment.
Inspect the water bath and heater / circulator tubing for algae or other materials. If
algae is present, change the bath water and add an algaecide. If you use a water
bath algaecide or clear bath product, ensure it is compatible with PETG and
acrylic.
Monthly Maintenance
See “Paddle / Basket Shaft Care” on page 99, “Basket Care” on page 100, and “Water
Bath / Acrylic Care” on page 101 as applicable for additional information on proper
maintenance of your equipment.
Drain the water from the water bath and clean the bath thoroughly. Refill the water
bath and add an algaecide.
Clean and lubricate the upper and lower portion of the support legs with the
recommended synthetic lubricant.
Caution
Use care when washing the TruCenter vessels as heat over 65 °C
can damage the plastic of the magnetic ring flanges.