EasyManuals Logo

Ametek 933S User Manual

Ametek 933S
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #242 background imageLoading...
Page #242 background image
6-66 | Model 933S UV Analyzer
Decimal Alarm
Bit Value Type Alarm Condition / Description and Suggested Corrective Action
0 1 Fault f: HC-OBMC communication problem
This serious system Fault condition indicates the Host Controller board cannot
establish reliable communication with the Microcontroller board.
Corrective Action:
Take appropriate safety precautions and cycle the analyzer AC power o󰀨 for a
minimum of 15 seconds before cycling it on again.
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Lower Enclosure, and:
From the Info dialog box, click the HC Reset button to reset the Host
Controller board.
If this still doesn’t correct the alarm, open the Electronics Enclosures and:
- Reset the Host Controller board (press SW100, Lower Enclosure – see
Figure 6-17).
- Reset the Microcontroller board (press SW400, Upper Enclosure – see
Figure 6-18).
Check the connections of the cable between the Host Controller (J300) and
Micro-Interface (J104) boards. Inspect it for damage (cuts, nicks, burn marks,
etc.).
Replace the Microcontroller, Host Controller, and/or Micro-Interface
boards. Contact AMETEK to verify operation before removing any of these
boards.
1 2 Fault f: Analytical data not received
This serious system Fault condition indicates the Host Controller board is not
receiving analytical data from the Microcontroller board.
Corrective Action:
Take appropriate safety precautions and cycle the analyzer AC power o󰀨 for a
minimum of 15 seconds before cycling it on again.
If this doesn’t correct the alarm, take appropriate safety precautions, open the
Lower Enclosure, and:
From the Info dialog box, click the HC Reset button to reset the Host
Controller board.
If this still doesn’t correct the alarm, open the Electronics Enclosures and:
- Reset the Host Controller board (press SW100, Lower Enclosure – see
Figure 6-17).
- Reset the Microcontroller board (press SW400, Upper Enclosure – see
Figure 6-18).
Check the connections of the cable between the Host Controller (J300) and
Micro-Interface (J104) boards. Inspect it for damage (cuts, nicks, burn marks,
etc.).
Replace the Microcontroller, Host Controller, and/or Micro-Interface
boards. Contact AMETEK to verify operation before removing any of these
boards.
Setup (tab)Analyzer
Info

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 933S and is the answer not in the manual?

Ametek 933S Specifications

General IconGeneral
BrandAmetek
Model933S
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals