4-51
4.4.10 Frequency Measurement via Input2 (10 Hz to 10 MHz)
To measure frequencies between 10 Hz and 10 MHz, select the Input2 connector and an impedance of 1M .
To measure frequencies between 10M Hz and 1 GHz, select the Input2 connector and an impedance of 50 . For more
details, see 4.4.9.
(1) Connecting the input signal
Connect the signal to measure to the Input2 connector on the front panel.
Note:
Do not connect a signal of 10 Vrms (at 1M )/2Vrms (at 50 ) or higher to the Input2 connector.
(2) Setup
1) Press the [Preset] key to preset this unit.
2) Set the input channel to Input2.
Press the [Input] key to display the input parameter setup screen. Enclosing the menu F1 in a thick frame with the
[<] and [>] keys and then pressing the [Enter] key select Input2.
3) Select an input impedance.
Enclosing the menu F2 in a thick frame with the [<] and [>] keys and then pressing the [Enter] key select 1M .
4) Switch between input attenuators.
When the input level of the signal to measure is low, turn off the attenuator. Enclosing the menu F3 in a thick
frame with the [<] and [>] keys and then pressing the [Enter] key select Off.
5) Pressing the [Return to Meas] key displays the normal measurement screen.
6) Using the [<] and [>] keys, set a desired frequency measurement resolution.
7) Using the [ ] and [ ] keys, set a desired sample rate.