4.3.17 Data Storage Function
This function is effective only when the unit is controlled through the GPIB.
After execution of a data storage start command, 100 pieces of frequency measurement data are stored in the internal
memory. When the 101st data is stored, the first data is invalidated, validating the second to 101st data (a total of 100
pieces of data). 100 pieces of data stored in the internal memory are updated until a data storage stop command is executed.
A stored data read command is executed to read the stored data.
0 Hz (execution error) is output in the following cases:
• When a stored data read command is executed without executing a data storage stop command after executing a data
storage start command
• When a data storage stop command or stored data read command is executed before 100 pieces of data have been stored
For details on the data storage start command, data storage stop command, and stored data read command, see Section 5,
“GPIB.”