Section 5 GPIB
5-10
5.4.4 Device Message List
(1) A
1) ACF frequency acquisition
Sets whether to acquire frequency manually or automatically, use the frequency value as value (manual fre-
quency) to measure during manual acquisition, or use the frequency value set by the manual frequency setting
command AF.
Command : ACF n (, s)
Query : ACF?
Response : ACF n
<Program data>
Value of n Value set
0 .......................... AUTO (initial value)
1 .......................... MANUAL
Value of s
0 .......................... Measures at the frequency set by command AF. (default value)
1 .......................... Measures at the frequency measured previously.
(AF setting is overwritten.)
2) ACL level acquisition
Set whether to acquire the level manually or automatically, use the current setting as the amplitude discrimination
value during manual acquisition, or use the frequency value set by the manual frequency setting command AF.
Command : ACL n (, s)
Query : ACL?
Response : ACL n
<Program data>
Value of n Value set
0 .......................... AUTO (initial value)
1 .......................... MANUAL
Value of s
0 .......................... Measures at the frequency set by command AD. (default value)
1 .......................... Measures at the level measured previously.
(AD setting is overwritten.)