4-43
4.4.5 Burst Wave Measurement via Input1
(Frequency Acquisition Mode = Manual; Level Acquisition Mode = Auto)
Setting Measurement Mode to Burst allows you to measure the carrier frequency, pulse width, and pulse repetition period
of the pulse-modulated signal. Perform steps (1) and (2) below.
(1) Connecting the input signal
Connect the signal to measure to the Input1 connector on the front panel.
Note:
Do not connect a signal of +10 dBm or higher to the Input1 connector.
(2) Setup
1) Press the [Preset] key to preset this unit.
2) Select the Burst mode.
Press the [Meas Mode] key. The Burst LED lights up.
3) For how to set the manual frequency value, see Section 4.4.2. Note that the input tolerance for burst measurement
is different from that for continuous wave measurement.
When the manual frequency value is within the range of 600 MHz to 1 GHz, the input tolerance is ±30 MHz.
When the manual frequency value is higher than 1 GHz and the narrow burst width is selected, the input tolerance
is ±20 MHz. When the manual frequency value is higher than 1 GHz and the wide burst width is selected, the
input tolerance is ±40 MHz.
4) To measure the pulse width and pulse repetition period at the same time, follow the procedure discussed in
Section 4.4.7.
Note:
If the frequency is not displayed at all or just not displayed properly, set Level Acquisition to Manual and then
perform measurements.